Pulse Oximeter

Category:

Description

  • 7 segment LED for SpO2, Pulse rate and for PI
  • 2.5″ LCD for SpO2 Graph (Plethysmograph)
Measurement of PI
  • Measure range : 0 ~ 20%
  • When the PI measure range is 1 ~ 20%, the permission of absolute error is ±1%
  • When the PI measure range is 0 ~ 0.9%, the permission of absolute error is ±0.2%
  • Resolution SpO2 ±1%; Pulse Rate: 1bpm; PI; 0.1%
  • Measurement Performance in weak filling condition; Sp02 and Pulse Rate value can be shown correctly when pulse-filling ratio is 0.4%
  • Sp02 error is ±1%. Pulse rate error is ±2bpm or ±2% [select larger]
Facebook
Twitter
WhatsApp
Product Details
Measurement of SpO2
  • Measure Range : 0 ~ 100%
  • Accuracy : When the Sp02 Measure range is 70 ~ 100%, the permission of absolute error is ±2%
  • Below 70% unspecified
Measurement of Pulse Rate
  • Measure Range: 30 ~ 250bpm
  • Accuracy: ±2bpm or ±2% [Select larger]
Operating Voltage
  • Internal power supply: 3.6V DC ~ 4.2V DC
  • Exterior alternating current supply: 100 ~ 240V AC, 50/60 Hz
  • OPTIONAL: NELLCOR technology for SpO2 Software
Resistance to surrounding light
  • The deviation between the value measured in the condition of man-made light or indoor natural light and that of darkroom is less than ±1%